Protocol A - 8-byte 55 AA command/response

Used by 14 models: LR1000E2/E4, LR1200E2/E4, LR1500E2/E4, LR2000E2/E4, LRF1200A1, LRF3000A1, SPD1200N0/N2/N4, SPD1200D03.

Every row below ends with the manual sentence it came from. <MODEL>.txt refers to that model's own manual transcription; the wording is identical across the eight LR-series manuals unless a row says otherwise.

FieldValueSource
InterfaceUART-TTL, 3.3 V"Communication mode: using serial communication mode" - LRF3000A1
Default baud115200"Source default baud rate: 115200. Data bits: 8. Frame length: 8 bytes." - LR1000E2
Data bits / stop / parity8 / (not stated) / (not stated)"Data bits: 8" - LR1000E2. Stop bits and parity are not stated in the family-A protocol sections.
Frame length8 bytes, both directions"Frame length: 8 bytes" - LR1000E2
Radixhexadecimal"All transmitted and received data is hexadecimal." - LR1000E2
Baud (SPD1200D03)conflict, unresolved"The PDF protocol section states 115200 bps as default, while the technical table states 15200 bps (factory) / 9600 bps." and "Do not select a host baud rate until the delivered unit configuration is confirmed." - SPD1200D03

Frame layout​

byte 0 1 2 3 4 5 6 7
55 AA Function D1 D2 D3 D4 Checksum

Source: "Frame Header (H) | Frame Header (L) | Function Code | D1 | D2 | D3 | D4 | Checksum" - LR1000E2 source row 1.

Checksums - the two are different, and the manual says so​

DirectionFormulaSource
Host → moduleSUM(Function + D1 + D2 + D3 + D4) & 0xFF"Source transmit checksum formula: SUM(Function Code + DATA1 + DATA2 + DATA3 + DATA4), lower 8 bits." - LR1000E2
Module → hostSUM(0x55 + 0xAA + Function + D1 + D2 + D3 + D4) & 0xFF"Source reply checksum formula: SUM(Frame Header (H) + Frame Header (L) + Function Code + DATA1 + DATA2 + DATA3 + DATA4), lower 8 bits." - LR1000E2

"The source warns that the transmit and response calculations differ." - LR1000E2

"The verification content for sending and receiving may differ, so please pay attention to discrimination." - LRF1200A1, LRF3000A1

The two scopes differ by exactly 0x55 + 0xAA = 0xFF, so a driver that uses the transmit formula on replies will be off by one everywhere and appear to work only when the checksum happens to wrap. This is the most likely family-A integration bug; python/tests/test_protocol_a.py pins both formulas.

Commands​

FunctionDirectionFrameSource
0x88 single-shotsend55 AA 88 FF FF FF FF 84"Single-Shot Ranging | Send | 55 AA 88 FF FF FF FF 84" - LR1000E2 row 7
0x88 single-shotreply55 AA 88 STA FF DIS_H DIS_L SUM"Single-Shot Ranging | Reply | 55 | AA | 88 | STA | FF | DIS_H | DIS_L | SUM[1:7]" - row 8
0x89 continuoussend55 AA 89 FF FF FF FF 85"Continuous Ranging | Send | 55 AA 89 FF FF FF FF 85" - row 11
0x89 continuousreply55 AA 89 STA FF DIS_H DIS_L SUMrow 12 - but see the protocol hold below for LRF1200A1 / LRF3000A1
0x8E stopsend55 AA 8E FF FF FF FF 8A"Stop Ranging | Send | 55 AA 8E FF FF FF FF 8A" - row 15
0x8E stopreply55 AA 8E STA FF FF FF SUM"PDF p8 / stop reply → 55 AA 8E STA FF FF FF SUM[1:7]" - SPD1200D03 (the LR-series row prints six cells; see Known gaps)
0x8A anglesend55 AA 8A FF FF FF FF 86"Angle Measurement | Send | 55 AA 8A FF FF FF FF 86" - row 19
0x8A anglereply55 AA 8A STA FF ANG_H ANG_L SUMrow 20. "Effective only for modules equipped with an angle sensor" - row 21
0x80 self-testreply only55 AA 80 STA 00 [00] ErrCode SUM"Self-Test Information | Reply | 55 | AA | 80 | STA | 00 | ErrCode | SUM[1:7]" - row 24; SPD1200D03 prints the eight-byte form
0x86 LD constant-ONsend55 AA 86 FF FF FF 01 84 (on) / 55 AA 86 FF FF FF 00 83 (off)row 33. Caution: "Do not keep the constant-ON mode enabled for long periods." Documented for the LR series and SPD1200N series only.
0x70 external circuit enablesend55 AA 70 AB CD 00 00 E8"External circuit enable | Send | 55 AA 70 AB CD 00 00 E8" - LRF3000A1 T5/R6. LRF1200A1 / LRF3000A1 only.
0x71 external circuit disablesend55 AA 71 AB CD 00 00 E9LRF3000A1 T5/R10. Puts the module into the manual's "secondary low-power mode".
baud setsend55 AA TYPE FF FF FF FF SUM"Baud Rate | Send | 55 | AA | TYPE | FF | SUM[3:7]" - row 28
baud set (SPD1200D03 only)send55 AA 01 FF FF FF TYPE SUM"PDF p8 / baud request → 55 AA 01 FF FF FF TYPE SUM[3:7]" - SPD1200D03

Status and result fields​

FieldMeaningSource
STA1 = success, anything else = failure"STA = other value: Measurement failed; STA = 1: Measurement successful" - LR1000E2 row 9
DIS_H, DIS_Ldistance, big-endian (DIS_H << 8 | DIS_L)"DIS_H: High byte of the measurement result; DIS_L: Low byte of the measurement result" - row 9
scalingraw counts are the value × 10"all measurement results are output as the actual value × 10" - row 9. Not stated for SPD1200D03.
ANG_H, ANG_Langle, same layout and ×10 scalingrow 21
ErrCodesee the table belowLRF1200A1 / LRF3000A1 T5/R15–R18

Error codes (LRF1200A1 / LRF3000A1)​

CodeDescription (verbatim)
0x00No echo signal was received
0x16Out of range: below the minimum range
0x18No echo signal was received
0x00–0x07Hardware error

The manual lists 0x00 twice - once as "no echo" and once inside the hardware error range. Reproduced as printed.

Baud-rate TYPE tables​

Two different tables exist. Do not merge them.

TYPELR / LRF-A1 / SPD1200NSPD1200D03
0196009600
021440014400
031920019200
043840038400
055600057600
0657600115200
07115200230400
08128000-
09230400-

Sources: "TYPE = 01: Set baud rate to 9600 bps … TYPE = 09: Set baud rate to 230400 bps" - LR1000E2 row 29; "TYPE 01 = 9600; 02 = 14400; 03 = 19200; 04 = 38400; 05 = 57600; 06 = 115200; 07 = 230400 bps. Restart required." - SPD1200D03.

Both manuals add that the change is not immediate: "The baud rate change does not take effect immediately; it becomes effective only after a reboot."

PROTOCOL HOLD - continuous ranging on LRF1200A1 and LRF3000A1​

"The source-stated continuous-ranging request uses function 0x89 while its recorded reply uses 0x88. This document preserves both and does not authorize continuous operation or a response parser."

"PROTOCOL HOLD Continuous ranging remains an engineering-confirmation item because the source request and reply function bytes differ."

Both drivers in this repository refuse continuous ranging on these two models unless the caller explicitly acknowledges the hold (stream(acknowledge_protocol_hold=True) in Python, startContinuous(true) in Arduino). Single-shot ranging is unaffected.

Known gaps​

GapWhat the manual does or does not say
Stop bits and parityThe family-A protocol sections state "Data bits: 8" and nothing about stop bits or parity. This repository does not set them and leaves the pyserial/Arduino defaults (8N1) in place; confirm with ERDI if your unit disagrees.
SPD1200D03 default baudIts own manual states 115200 bps in the protocol section and 15200/9600 bps in the technical table, and instructs the reader not to choose until the delivered unit is confirmed. The drivers therefore require an explicit baud rate for this model.
SPD1200D03 distance scalingNo scaling statement of any kind appears. The Python driver returns raw counts and metres = None; the Arduino driver returns NAN. Nothing is inferred from the sibling SPD1200N models.
SPD1200D03 STA semanticsNot stated, so the status byte is returned unchecked rather than judged against 1.
SPD1200N0/N2/N4 transmit checksumThe transcription says request and reply scopes differ but does not print the transmit formula. The drivers use the family-A transmit formula, which reproduces every example frame these manuals do print (55 AA 88 FF FF FF FF 84, … 89 … 85, … 8E … 8A, … 8A … 86). Flagged rather than assumed silently.
Self-test ErrCode byte positionThe LR-series row prints seven cells (55 | AA | 80 | STA | 00 | ErrCode | SUM), putting ErrCode at D3; SPD1200D03 prints the full eight bytes with ErrCode at D4. The Python driver exposes both candidates (error_code_d3, error_code_d4) instead of picking one.
Stop-reply widthThe LR-series stop reply row prints six cells; only SPD1200D03 prints the padded eight-byte form. Treated as an eight-byte frame with the row's merged cells expanded.
LRF1200A1 / LRF3000A1 continuousSee PROTOCOL HOLD above.
Distance unitThe protocol sections state only the ×10 scaling, never the unit itself. Metres comes from each model's own ranging-range row in its source technical table.